Position pump in a way so that the disassembled pump

head is above, support if necessary.

By turning eccentric bushing (front of connecting rod),

bring connecting rod into a position in which diaphragm is

in contact with housing and centred with respect to bore.

Reassemble in reverse order.

Install head cover with O-ring, valves and housing cover

with housing cover insert.

Make sure that the valves are correctly seated: Valves at

the outlet with round centred opening under valve, valves

at the inlet with kidney-shaped opening beside valve.

Ensure that the diaphragm is positioned centrally so that it

will become clamped uniformly between housing and head

cover.

Assembling the pump head

Screw in four socket head screws fixing housing cover

crosswise (e. g. in the sequence 1,2,3,4) first slightly, then

tighten.

Do not tighten until head cover is in contact with housing,

torque 12 Nm.

Attention: Ensure that the diaphragm clamping disc

never hits the head cover! In case of loud knocking

switch off the pump immediately.

Inspect pump or replace diaphragms and valves at the

other pump heads analogue.

3

2

1

4

Measure the pressure at the inlet port of the individual

head: Use a suitable vacuum gauge (e. g. DVR 2, cat.

no.: 682902), make sure that it is correctly calibrated, and

measure the pressure at the inlet port. A vacuum of less

than 120 mbar should be indicated.

If the reading is higher, recheck the pump chamber and

make sure that the valves and the diaphragms are cor-

rectly seated (diaphragms concentric with bore).

Screw housing plate to pump.