ADLINK PXI-2006 User Manual

Page 39

Advertising
background image

Operation Theory

29

due to the variation of the conver-sion time of the A/D con-

verters.

Table 4-1 and 4-2 illustrate the ideal transfer characteristics of var-
ious input ranges of DAQ\PXI-20XX. The converted digital codes
for DAQ\PXI-2010 are 14-bit and 2’s complement, and here we
present the codes as hexa-decimal numbers. Note that the last 2
bits of the transferred data, which are the synchronous digital input
(SDI), should be ignored when retrieving the analog data.

Description

Bipolar Analog Input Range

Digital code

Full-scale Range

±10V

±5V

±2.5V

±1.25V

Least significant bit 1.22mV 0.61mV 0.305mV 0.153mV

FSR-1LSB

9.9988V 4.9994V 2.4997V

1.2499V

1FFF

Midscale +1LSB

1.22mV 0.61mV 0.305mV 0.153mV

0001

Midscale 0V

0V

0V

0V

0000

Midscale –1LSB

-1.22mV -0.61mV -0.305mV -0.153mV

3FFF

-FSR

-10V

-5V

-2.5V

-1.25V

2000

Table 4-1: Bipolar analog input range and the output digital code on DAQ/

PXI-2010 (Note that the last 2 digital codes are SDI<1..0>)

Description

Unipolar Analog Input Range

Digital code

Full-scale Range

0V to 10V 0 to +5V 0 to +2.5V 0 to +1.25V

Least significant bit

0.61mV

0.305mV

0.153mV

76.3uV

FSR-1LSB

9.9994V

4.9997V

2.9999V

1.2499V

1FFF

Midscale +1LSB

5.00061V 2.50031V 1.25015V

625.08mV

0001

Midscale 5V

2.5V

1.25V

625mV

0000

Midscale –1LSB

4.99939V 2.49970V 1.24985V

624.92mV

3FFF

-FSR

0V

0V

0V

0V

2000

Table 4-2: Unipolar analog input range and the output digital code on DAQ/

PXI-2010 (Note that the last 2 digital codes are SDI<1..0>)

Advertising
This manual is related to the following products: