Test box high distortion or noise, Test box curves inconsistent, Test box curves differ from specifications – Audioscan Axiom User Manual
Page 100: Test box speaker overdriven, 20 troubleshooting, Self test failures, Initialize function

Axiom
®
User's Guide Version 1.8
May 2015
20 Troubleshooting
This section covers common problems encountered when using the Axiom. If you require further assistance,
note your model and serial number (located just behind the test box lid) and contact your local distributor, or
Audioscan at 519-268-3313 800-265-2093 (USA only), 519-268-3256 (fax) or [email protected]
PLEASE DO NOT SHIP YOUR UNIT BACK TO THE FACTORY. In most cases we can courier you a part
that you can quickly and easily replace yourself.
Self test failures
Any failure during power-on self test (P.O.S.T.) is indicated on-screen and further use of the instrument is
denied. Try the following steps to resolve the failure:
1. To access Self test, click on
and then click on [Self test].
2. If you received a Max out/in or a Routing failure, remove any hearing aids from the test chamber and rerun
self test.
3. Unplug all microphones (On-ear & Test box) from the test chamber. Rerun self test. If self test passes,
reconnect the microphones, one by one, running self test each time until a failure occurs. If the failure
returns, display and print the test results and contact Audioscan for help.
Initialize Function
Initializing your Audioscan unit will reset it to default factory settings. This erases the user preference files
stored in internal memory. Any custom printer choice, network settings, test setups will need to be re-entered
and the user calibrations will also be required.
1. To initialize, click on [Setup], then click on [Self test]
2. Click on [Initilize]. Unit will automatically restart.
100