Fm fitting and verification – Audioscan Axiom User Manual

Page 94

Advertising
background image

Axiom

®

User's Guide Version 1.8

May 2015

To test frequency-lowering:

1. Click on

and choose Speechmap from either the On-ear or Test box menus.

2. Program the hearing instrument so that the frequency-lowering feature is inactive.

3. Access Test 1, 2 or 3, then, click on

to start that test.

4. Choose Speech3150, Speech4000, Speech5000 or Speech6300 from the Stimulus pull-down menu. Choose

65dBSPL from the Level pull-down menu. Note that only the LTASS is shown for these stimuli and targets
(if selected) have been suppressed.

5. Click on

to measure and store the long-term signal-averaged result. If the isolated band does not appear

in the Test 1 curve, it may indicate that the hearing aid has no gain at the selected band frequency. This does
not mean that the hearing aid will be unable to transpose the band to a lower frequency when frequency
lowering is enabled.

6. Program the hearing instrument to activate the frequency-lowering feature.

94