17 speechmap, Speechmap facts, Dsl 5.0 in speechmap – Audioscan Axiom User Manual
Page 76
Advertising

Axiom
®
User's Guide Version 1.8
May 2015
5. Set Instrument to [CROS].
6. Select REAR 2 and obtain a response using [Pink noise] at 55. Adjust the response to match that obtained in
step 3.
7. Position the speaker directly in front of the client. Select REAR 3 and use [Pink noise] at 55 to check for a
smooth response. Repeat with Instrument set to [BTE]. An irregular response may be the result of reflections
from nearby objects (including the operator) or it may indicate phasing problems in the aids.
76