Bronkhorst High-Tech B.V.

IQ+FLOW

9.17.045

12

3

Basic operation

3.1

General

An IQ

+

FLOW® instrument must be powered with +15…+24 Vdc according to the applicable hook-up diagram, supplied with the

instrument. The instrument can be operated by means of:

Analog interface: 0…5 Vdc; 0…10 Vdc; 0…20 mA or 4…20 mA (single-channel versions only)
Digital RS232 interface (FLOW-BUS (Propar) protocol)
Digital RS485 interface (Modbus RTU, Modbus ASCII or FLOW-BUS protocols)

By default the instrument is set as specified. The table below lists the supported interfaces for single-channel and multi-channel

versions.

Analog interface

(section 3.2)

Digital RS232 interface

(section 3.4)

Digital RS485 interface

(section 3.5)

IQF/IQP

(single-

channel)

0…5 Vdc; 0…10 Vdc;

0…20 mA; 4…20 mA

(software selectable)

FLOW-BUS (Propar) protocol on 9600,

19200, 38400, 57600 or 115200 Baud

(software selectable)

Modbus RTU and Modbus ASCII protocols

on 9600, 19200, 38400, 56000, 57600 or

115200 Baud; FLOW-BUS protocol on

187500 or 400000 Baud (software

selectable)

IQM

(multi-

channel)

Not supported

FLOW-BUS (Propar) protocol on 38400 or

115200 Baud (select with rotary switch)

Modbus RTU protocol on 9600, 19200 or

38400 Baud (select with rotary switch)

3.2

Analog operation

Analog operation is only possible with single-channel IQ

+

FLOW® instruments. Multi-channel instruments can only be operated

digitally. At analog operation the following signals are available:

Measured value (analog output) at pin 2
Setpoint (analog input/setpoint) at pin 3

The valve output is no longer available as analog signal with IQ

+

FLOW®. The selected analog interface (0…5 Vdc; 0…10 Vdc; 0…20

mA or 4…20 mA) can be found in the model key of the instrument. Refer to Section 2.13.

When operating the instrument through the analog interface it is possible to connect the instrument simultaneously

to RS232 for reading/changing parameters (e.g. settings or fluid selection).

Hook-up

For analog operation either an RJ-45 loose-end cable or an RJ-45 to 9-pin sub-D converter may be used to connect the required

signals.

Refer to the “Hook-up diagram IQ

+

FLOW®” or use an

RJ-45 loose-end cable (7.03.419) to connect the required

signals.

When using a Bronkhorst readout unit use only an RJ-45

cable (7.03.236) in combination with the RJ-45 to 9-pin

sub-D converter (7.03.376). With these items the pin

configuration is unchanged.