Compression type fittings

For leak tight installation of compression type fittings make sure that the tube is inserted up to the shoulder in the

fitting body and that no dirt or dust is present on tube, ferrules or fittings. Tighten the nut finger tight; while holding

the instrument, then tighten the nut one turn.

10-32 UNF fittings

Tighten the 10-32 UNF fittings according to the instructions of the

supplier of the fittings.

Only use 1/16” tubing with a straight and clean cut without burrs to

ensure leak tightness. Preferably deburr the tubing prior to

installation. A new ferrule connection must be made for each new

adapter to ensure leak-tightness and minimum dead volume, due to

variances in the adapter dimensions.

Mounting downported instruments

For downported instruments make sure that the seals are present at the bottom, that the surfaces are undamaged

and that they are dry and free from dirt or dust.

2.5

Leak check

Check the system for leaks before applying (fluid) pressure. Especially if toxic or other dangerous fluids are used!

2.6

Electrical connection

Electrical connections must be made with a standard cable or according to the IQ

+

FLOW® hook-up diagram. Several

hook-up examples and standard cables are found in section 3.
IQ

+

FLOW® instruments are powered with +15…+24 Vdc.

i

www

Electrical connections must be made with standard cables or according to the applicable hook-up diagrams. These

documents can be found at: http://www.bronkhorst.com/en/downloads

The instruments contain electronic components that are susceptible to damage by electrostatic discharge. Proper

handling procedures must be taken during installation, removing and connecting the electronics.

The instruments described in this manual carry the CE-mark and are complient with the EMC requirements.

However compliance with the EMC requirements is not possible without the use of proper cables and connector/

gland assemblies. Bronkhorst High-Tech B.V. recommends the use of their standard cables. These cables have the

right connectors and if loose ends are used, these will be marked to prevent wrong connection. When using other

cables, cable wire diameters should be sufficient to carry the supply current and voltage losses must be kept as low

as possible. When in doubt: contact your distributor.

When connecting the system to other devices (e.g. to PLC), be sure that the integrity of the shielding is not affected.

Do not use unshielded wire terminals.

2.7

Analog / digital operation

Analog operation (single channel versions only)

For analog operation refer to the “Hook-up diagram IQ

+

FLOW®” or use an RJ-45

loose-end cable (7.03.419) to connect the required signals.