Advanced measurement and control parameters

Measured Value (Fmeasure)

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Float

R

-3.4E+38…

3.4E+38

205

33/0

0xA100…0xA101/41217…41218

Floating point variable of the 'Measured Value'. The 'Fmeasure' variable shows the measured value in the capacity and capacity unit

for which the instrument has been set. The 'Fmeasure' parameter is dependent of 'Capacity', 'Capacity Unit', 'Sensor Type' and

'Capacity 0%'.

Setpoint (Fsetpoint)

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Float

RW

0…3.4E+38

206

33/1

0xA119…0xA11A/41241…41242

Floating point variable of the 'Setpoint'. The 'Fsetpoint' variable shows the setpoint value in the capacity and capacity unit for which

the instrument has been set. The 'Fsetpoint' parameter is dependent of 'Capacity', 'Capacity Unit', 'Sensor Type' and 'Capacity 0%'.

Valve Output

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned long

RW

0…

16777215

55

114/1

0x001F/32

This parameter is the signal coming out of the controller, used for driving the valve. 0…16777215 corresponds with approximately

0…60 mAdc.

4.1.4

Controller parameters

The controlling algorithm for the valve handled by the micro-controller consists of several parameters which can be set.

Although many parameters could be accessed, Bronkhorst® advises not to change these parameters because during manufacturing

the controller is optimized. Changing of controller settings should be performed by or under supervision from trained service

personnel only.

The picture below shows the basic controller diagram of the digital instrument. It consists of a standard PID controller with a

number of add-ons. Basically, when a faster or slower controller response is needed, only the controller speed (Kspeed) or PID-Kp

should be changed.

Kp (PID-Kp)

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Float

RW

0…1E+10

167

114/21

0xF2A8…0xF2A9/62121…62122

PID controller proportional action, multiplication factor.