IQ+FLOW

25

9.17.045

Dynamic Display Factor

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Float

RW

0…1.0

56

117/1

0xF508…0xF509/62729…62730

This value should not be changed.

Static Display Factor

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Float

RW

0…1.0

57

117/2

0xF510…0xF511/62737…62738

This value should not be changed.

CycleTime

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned char

R

0…255

52

114/12

0x0E4C/3661

Note: The unit of parameter CycleTime is 10 ms. Example: value 0.2 means 2 ms

4.1.6

Alarm / Status parameters

i

www

See document nr. 9.17.023: Operational instructions for digital instruments, chapter 6.

http://www.bronkhorst.com/files/downloads/

manuals_english/917023_operation_instructions_digital_instruments.pdf

4.1.7

Counter parameters

i

www

See document nr. 9.17.023: Operational instructions for digital instruments, chapter 7.

http://www.bronkhorst.com/files/downloads/

manuals_english/917023_operation_instructions_digital_instruments.pdf

4.2

Special instrument features

4.2.1

Auto zeroing

To start the auto zero-procedure by digital operation two parameters should be written:

Control Mode

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned char

RW

0…255

12

1/4

0x0024/37

Calibration Mode

Type

Access

Range

FlowDDE

FLOW-BUS

Modbus

Unsigned char

RW

0…255

58

115/1

0x0E61/3682

Value

Mode

Instrument action

0

Idle

Idle

9

Auto zero

Auto-zeroing

255

Error

Idle