2 led functions, Led functions, Section 3.3.2 – Bronkhorst IQ+FLOW (from 01-07-2013) User Manual

Page 14: Section 3.3.3

Advertising
background image

Bronkhorst High-Tech B.V.

IQ+FLOW

9.17.045

14

LEDs

Time

Pushed

Indication

Green

Red

Off

Off

0 – 4 sec.

No action. Pressing a switch shortly by accident will not start any unwanted

reaction of the instrument.

Off

Normal

flash

0,2 sec on,

0,2 sec off

4 – 8 sec.

Restore factory settings

All parameter settings (except field bus settings) will be restored to situation of

final test at Bronkhorst production.

Normal

flash

0,2 sec on,

0,2 sec off

On

8 – 12 sec.

For FLOW-BUS only: install a free node-address on FLOW-BUS.

Normal

flash

0,2 sec on,

0,2 sec off

Normal

flash

0,2 sec on,

0,2 sec off

12 – 16 sec

Activate 'Configuration Mode'. The baud rate and bus type are set to 38k4 and

RS232 FLOW-BUS (Propar). The 'Configuration Mode' is recognized by the

green LED blinking 2 sec on, 0,1 sec off. The 'Configuration Mode' is

deactivated only after applying this micro-switch action again.

LED indications using micro-switch at power-up situation of an instrument

3.3.2

LED functions

The LEDs on top of the instrument can also be used for manual operation of some options. The green LED will indicate in what

mode the instrument is active. The red LED will indicate error/warning situations.

i

www

For details see “Manual interface: micro-switch and LEDs” in Operation Instructions Digital Instruments:

http://www.bronkhorst.com/files/downloads/

manuals_english/917023_operation_instructions_digital_instruments.pdf (document nr. 9.17.023, Chapter 11)

3.3.3

Rotary switch operation (multi-channel versions only)

The IQ

+

FLOW® multi-channel instruments are equipped with rotary switches for selection of communication type, baud rate and

node address.

Communication type switch

With the 'COMM. TYPE' switch the following communication types can be selected:

0. RS485 9k6: Modbus RTU protocol with baud rate 9600, EVEN parity

1. RS485 19k2: Modbus RTU protocol with baud rate 19200, EVEN parity

2. RS485 38k4: Modbus RTU protocol with baud rate 38400, EVEN partiy

3. RS232 38k4: FLOW-BUS protocol with baud rate 38400

4. RS232 115k2: FLOW-BUS protocol with baud rate 115200

Node address switches

With the two 'NODE ADDRESS' switches the node address of the instrument channels can be selected. The 'MSD' (Most Significant

Digit) sets the first digit (tens), the 'LSD' (Least Significant Digit) sets the second digit (units). The node address of channel 1 is set

with the switches, the channels 2 and 3 receive 'node address' + 1 and 'node address' + 2 respectively (e.g. MSD = 1 and LSD = 9

selects node 19 for channel 1, but also node 20 and 21 for channels 2 and 3).