Setup

Setting Communication Parameters

4

4-3

4.3 Setting Communication Parameters

From the Communications Menu, the user can set/edit the following parameters:

• Output Type (Factory Setting: Computer)
• Output Mode (Factory Setting: Manual)
• Baud Rate (Factory Setting: 9600)
• Parity (Factory Setting: None)
• Word Length (Factory Setting: 8 bits)
• Handshaking (Factory Setting: DSR Ignored)

















Figure 4-2. Communications Menu

Use the following procedure:

1. Press or to scroll through the Setup Menu labels until Communications is displayed.

2. Press

ENTER. The following message will be displayed:

Output

Type.

3. Follow the procedures outlined below to edit applicable setup parameters/values.

4. Exit the Communications Menu as follows:

a. Press or to scroll through the Setup Menu labels until Exit Com. Menu is displayed.

b. Press ENTER. The display will return to the Setup Menu and Clear Data Log will be displayed.

Output Type

The selected output type tells the Densitometer whether the RS-232 port is connected to a computer or a
printer:

1. Press or to scroll through the Communications Menu until Output Type is displayed.

2. Press

ENTER. The currently selected output type will be displayed.

3. Press or to scroll through the available output types (Printer or Computer).

4. When the output device you are using is displayed, press ENTER to make the selection. The

display will return to the Communications Menu and Output Mode will be displayed.

Enter

Enter

Enter

Enter

Enter

Enter

Enter

Enter

Enter

Enter

Enter

Enter

Communications

Output

Type

Output

Mode

Baud Rate

Parity

Word

Length

Hand-

shaking

Printer/

Computer

Manual/

Automatic

9600, 1200,

2400, 4800

None/Space

Mark/Even/

Odd

7 Bits/

8 Bits

DSR

Required/

DSR Ignored

Clear

Data Log

Exit

Comm.

Menu

Enter

Enter