Setup

D log E Setup

4

4-13

This procedure allows the user to setup, maintain, and selectively clear the control chart files. The setup
of these files includes the automatic (Scan New Params) or manual (Enter New Params) entry of the
following parameters in Tables 4-1 and 4-2.

Table 4-1.

Index Chart File

Chart Item

Range

Default
Setting

Contrast Index



Baseline
Deviation

0.00 - 4.00

0.02 - 3.98
(1)



1.25
0.15

Speed Index




Baseline
Deviation

0.00 - 4.00

0.02 - 3.98
(1)



1.00
0.15

Base + fog




Baseline
Deviation

0.00 - 4.00

0.02 - 3.98
(1)



0.20
0.02

Max. OD




Baseline
Deviation

0.00 - 4.00

0.02 - 3.98
(1)



3.00
0.15

Table 4-2.

Gradient Chart File

Chart Item

Range

Default
Setting

Average
Gradient



Baseline
Deviation

0.00 - 9.99

0.02 - 9.97
(1)



2.00
0.15

Relative speed




Baseline
Deviation

0.00 - 3.00

0.02 - 2.98
(1)



1.00
0.15

Base + fog




Baseline
Deviation

0.00 - 4.00

0.02 - 3.98
(1)



0.20
0.02

Max. OD




Baseline
Deviation

0.00 - 4.00

0.02 - 3.98
(1)



3.00
0.15




The deviation range is dependent on the baseline
value; the deviation must be less than half the
baseline value. This limit allows the control chart to
contain two deviation intervals above the baseline
and two below the baseline.

NOTE