Ecg simulation – Fluke Biomedical ProSim 4 User Manual

Page 18

Advertising
background image

ProSim™ 4

Users Manual

8

Table 7. Pre-Defined Patient Simulations

Simulation Name

Parameter Pre-Set

Value

Normal

Wave Form

NSR (Adult)

ECG Rate

60 bpm

Respiration Rate

20 bpm

IBP Channel 1

120/80 mmHg (Art)

NIBP

120/80 (93) mmHg

Hypertensive

Wave Form

NSR (Adult)

ECG Rate

120 bpm

Respiration Rate

40 bpm

IBP Channel 1

200/150 mmHg (Art)

NIBP 200/150

(167)

mmHg

Hypotensive

Wave Form

NSR (Adult)

ECG Rate

40 bpm

Respiration Rate

10 bpm

IBP Channel 1

60/0 mmHg (LV)

NIBP

60/30 (40) mmHg

Control of heart rate, respiration, invasive blood pressure, and non-invasive

blood pressure simulations are also accessed through the controls on page 1 of

the Home screen. A number of arrythmias are set through the Arrhythmia

control. Refer to the applicable section to learn more about each simulation

function.
To show page 2 of the Home screen, push the More softkey. Page 2 shows

controls that access autosequences, pressure and leak tests, and all setup

parameters. To go back to page 1 of the Home screen, push the Back softkey.

ECG Simulation

The Product simulates normal heart signals (ECG) as well as a variety of heart

arrhythmias. You set adult or neonatal simulation and change the heart rate

(beats per minute) through the ECG control.
To measure the ECG performance of a monitor, connect the Product to the

monitor as shown in Figure 2. A maximum of ten ECG leads can be connected to

the Product.