Arrhythmia simulation – Fluke Biomedical ProSim 4 User Manual

Page 20

Advertising
background image

ProSim™ 4

Users Manual

10

Heart rate, respiration, invasive blood pressure, and non-invasive blood pressure

parameters are shown at the top of the screen. To change the heart rate, touch

one of the pre-defined heart rate controls in the display.
The ECG function can be set to an adult or neonatal ECG signal. To change

between Adult and Neonatal, touch the softkey control in the lower-left corner of

the display. The softkey text will change to the one value not set. For example,

when Neonatal is shown in the softkey, an adult ECG signal is simulated.
Touch the Home softkey to go back to the Home screen.
Table 8 shows the percentage of the signal amplitude value that is put on each

ECG lead.

Table 8. ECG Lead Amplitudes

Waveform I

II

III

V1

V2 V3 V4 V5 V6

Normal Sinus & Performance

70 % 100 %

30 % 24 % 48 % 100 %

120 % 112 %

80 %

Arrhythmia Simulation

The Product can simulate a number of arrhythmias on the ECG leads. As well as

physiological waveforms, the Product can supply a signal to measure the

performance of an ECG monitor. To do an arrhythmia simulation or do a monitor

test with a performance wave, touch the Arrhythmia control in the Home screen.

Page 1 of the screens in Figure 4 shows in the display.

Page 1

Page 2

gne008.bmp

Figure 4. Arrhythmia Screens