Vital Signs Simulator

Autosequences

21

Push  to abort the step and move to the subsequent step. When the

autosequence has moved to step two,  shows in the display. Touch  to

pause the step.  shows in the display when the auto sequence is paused.

Push  to continue the step for the time left when the step was paused.
To abort the autosequence, touch the Stop softkey and go back to the

autosequence view screen.
Table 9 is a list of autosequences that are built into the Product.

Table 9. Autosequences

Autosequence

Sequence Steps

Run Time

Cardiac Failure

ECG 80 bpm, IBP Off, NIBP Off, and Respiration Off

00:45

ECG PVC (LV), IBP Off, NIBP Off, and Respiration Off

00:30

ECG VTach (LV) 160 bpm, IBP Off, NIBP Off, and Respiration Off

00:30

ECG VFib Coarse, IBP Off, NIBP Off, and Respiration Off

00:30

ECG Asystole, IBP Off, NIBP Off, and Respiration Off

00:15

STOP Total

Time

02:30

xercise

ECG 60 bpm, IBP Off, NIBP Off, and Respiration Off

00:30

ECG 90 bpm, IBP Off, NIBP Off, and Respiration Off

00:30

ECG 120 bpm, IBP Off, NIBP Off, and Respiration Off

00:30

ECG 150 bpm, IBP Off, NIBP Off, and Respiration Off

00:30

ECG 90 bpm, IBP Off, NIBP Off, and Respiration Off

00:30

ECG 60 bpm, IBP Off, NIBP Off, and Respiration Off

00:30

REPEAT Total

Time

03:00

Respiration

ECG 60 bpm, IBP Off, NIBP Off, and Respiration 110 brpm

00:30

ECG 60 bpm, IBP Off, NIBP Off, and Respiration 60 brpm

00:30

ECG 60 bpm, IBP Off, NIBP Off, and Respiration 20 brpm

00:30

ECG 60 bpm, IBP Off, NIBP Off, and Respiration 0 brpm (Apnea) 00:12

REPEAT Total

Time

01:42

Monitor Testing

ECG 120 bpm, IBP 200/150 (Art), NIBP 200/150 (167), and
Respiration 40 brpm

01:30

ECG 60 bpm, IBP 120/80 (Art), NIBP 120/80 (93), and Respiration
20 brpm

01:30

ECG 40 bpm, IBP 60/0 (LV), NIBP 60/30 (40), and Respiration
10 brpm

01:30

STOP Total

Time

04:30

Advertising