Detailed specifications, Normal-sinus-rhythm waveform, Arrhythmia – Fluke Biomedical ProSim 4 User Manual

Page 38

Advertising
background image

ProSim™ 4

Users Manual

28

Electromagnetic Compatibility (EMC)

International ........................................................ IEC 61326-1: Basic Electromagnetic Environment

CISPR 11: Group 1, Class A

Group 1: Equipment has intentionally generated and/or uses
conductively-coupled radio frequency energy that is necessary for
the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other
than domestic and those directly connected to a low-voltage power
supply network that supplies buildings used for domestic purposes.
There may be potential difficulties in ensuring electromagnetic
compatibility in other environments due to conducted and radiated
disturbances.
Emissions that exceed the levels required by CISPR 11 can occur
when the equipment is connected to a test object.

USA (FCC) .......................................................... 47 CFR 15 subpart B

Detailed Specifications

Normal-Sinus-Rhythm Waveform

ECG Reference ...................................................... The ECG amplitudes specified are for Lead II (calibration), from the

baseline to the peak of the R wave. All other leads are proportional.

Normal Sinus Rhythm .......................................... 12-lead configuration with independent outputs referenced to right leg

(RL). Output to 10 Universal ECG Jacks, color-coded to AHA and IEC
Standards.

Amplitude .............................................................. 1.0 mV. Other leads are proportional to Lead II (reference lead) in

percentage per:

Lead I .................................................................. 70
Lead II ................................................................. 100
Lead III ................................................................ 30
Lead V1 .............................................................. 24
Lead V2 .............................................................. 48
Lead V3 .............................................................. 100
Lead V4 .............................................................. 120
Lead V5 .............................................................. 112
Lead V6 .............................................................. 80

Amplitude Accuracy .............................................

±

5 % of setting Lead II

ECG Rate ............................................................... 30, 60, 80, 90, 120, 150, 180, 210, 240, 270, 300, and 320 BPM

(Preset Hypotensive condition is at 40 BPM)

Rate Accuracy .......................................................

±

1 % of setting

ECG Waveform Selection ..................................... Adult (80 ms) or neonatal (40 ms) QRS duration
Power-On Default .................................................. 60 BPM, 1.0 mV, adult QRS

Arrhythmia

Atrial Fibrillation ................................................... Coarse or fine
Premature Ventricular Contraction ..................... Left Ventricular
Ventricular Tachycardia ....................................... 160 or 200 BPM
Ventricular Fibrillation .......................................... Coarse or fine.
Transvenous Pacer Pulse .................................... 75 BPM, left arterial, 3 mV amplitude on lead II, Accuracy

±

10 %,

1.0 ms width

2nd Degree AV Block ............................................ Type 1
3rd Degree AV Block
Asystole

Advertising