Fluke Biomedical ProSim 4 User Manual

Page 26

Advertising
background image

ProSim™ 4

Users Manual

16

To do a leak test:
1. Touch the More softkey.
2. Touch the Leak Test and Manometer control to show the screen shown in

Figure 10 in the display.

gne011.bmp

Figure 10. Leak Test Screen

3. Touch  or  to set the test time for the leak test. The maximum test time is

5 minutes and the minimum test time is 30 seconds (default).

4. Manually increase the pressure in the pneumatic system to the pressure at

which the test is to start. When the measured pressure is 10 mmHg or higher,

the left softkey shows Start.

5. Touch the Start softkey. The timer counts down the test time.