Vital Signs Simulator

Detailed Specifications

29

ECG-Performance-Testing

Amplitude .............................................................. 1 mV. Other leads are proportional to Lead II (reference lead) in

percentage per:

Lead I .............................................................. 70
Lead II ............................................................. 100
Lead III ............................................................ 30
Lead V1 .......................................................... 24
Lead V2 .......................................................... 48
Lead V3 .......................................................... 100
Lead V4 .......................................................... 120
Lead V5 .......................................................... 112
Lead V6 .......................................................... 80

Square Wave ......................................................... 60 ms at 2.0 Hz

Respiration

Rate ........................................................................ 0 (OFF), 10 to 100 BrPM in 10 BrPM steps
Impedance Variations (

Δ

Ω

) ................................. 1

Ω

Accuracy Delta ......................................................

±

(10 % + 0.05 ohm)

Baseline ................................................................. 500

Ω

to circuit common, giving 1000

Ω

between any two leads

Accuracy Baseline ................................................

±

5 %

Respiration Lead ................................................... LA or LL (default)

Invasive Blood Pressure

Channels ................................................................ 1 electrically isolated from all other signals
BP Output
.............................................................. Circular DIN 5-Pin
Input/output Impedance ....................................... 300

Ω

±

10 %

Exciter Input Range .............................................. 2.0 to 16.0 V peak
Exciter-Input Frequency Range ........................... DC to 5000 Hz
Transducer Sensitivity ......................................... 5

μ

V/V/mmHg

Pressure Accuracy ...............................................

±

(1 % of setting + 1 mmHg) Accuracy guaranteed for DC excitation

only

Static Pressure ...................................................... 0, 80, 160, and 250 mmHg
Dynamic Waveforms

Synchronization .................................................. To ECG heartrate

Chambers simulated systolic/diastolic pressure:

Type

IBP (Arterial)

IBP (Left Ventrical)

Adult 60/30

60/0

Adult 120/80

120/0

Adult 150/100

150/0

Adult 200/150

200/0

Neonatal 35/15

35/0

Neonatal 70/40

70/0

Non-Invasive Blood Pressure

Pressure Units ....................................................... mmHg
Manometer (Pressure Meter)

Range ................................................................. 10 to 400 mmHg
Resolution ........................................................... 0.1 mmHg (for display purposes)
Accuracy .............................................................

±

(1 % reading +1 mmHg)

Pressure Source ................................................... Inflation bulb or device under test