How to set an arrhythmia ecg waveform, How to output a performance wave, Respiration simulation – Fluke Biomedical ProSim 4 User Manual

Page 21

Advertising
background image

Vital Signs Simulator

Respiration Simulation

11

How to Set an Arrhythmia ECG Waveform

Connect the Product to an ECG monitor as shown in Figure 2. The Product sets

the coarse atrial fibrillation waveform when the Arrhythmia screen shows in the

display. There are eight arrhythmia waveform controls shown in page 1 of the

Arrhythmia screen. Controls for more arrhythmia waveforms and a performance

wave are shown in page 2 when the More softkey is touched.
The simulated arrhythmia waveform and the respiration rate are shown at the top

of the display. The invasive and non-invasive blood pressure simulations are

disabled when an arrhythmia waveform is simulated. To change the arrhythmia

waveform, touch an arrhythmia control in page 1 or page 2 of the Arrhythmia

screens.
Touch the Home softkey to go back to the Home screen.

How to Output a Performance Wave

One performance wave can be output on the ECG leads for an ECG monitor test.

To output a square wave on the ECG leads:
1. Touch the Arrhythmia control in the Home screen.
2. Touch the More softkey to show page 2 of the Arrhythmia screens.
3. Touch the Performance: Square control in the display.
A square wave is output on the ECG leads until a control is touched or the Home

softkey is touched.
Touch the Home softkey to go back to the Home screen.

Respiration Simulation

The Product simulates respiration on the left leg or left arm ECG lead. To set a

respiration simulation parameter, touch the Respiration control in the Home

screen to show the screen in Figure 5 in the display.

gne004.bmp

Figure 5. Respiration Screen