Presets and autosequences – Fluke Biomedical ProSim 4 User Manual

Page 40

Advertising
background image

ProSim™ 4

Users Manual

30

NIBP Simulations

Pulse ................................................................... 2 mmHg max into 500 ml NIBP system
Volume of air moved ........................................... 1.0 ml max
Simulations ......................................................... Systolic/diastolic (MAP)

Adult ................................................................ 60/30 (40), 120/80 (93); 150/100 (117); and 200/150 (167)
Neonatal ......................................................... 35/15 (22) and 70/40 (50)

Repeatability ....................................................... Within

±

2 mmHg (at maximal pulse size independent of device under

test)

Synchronization .................................................. To ECG heartrate (maximal rate 120 BPM)

Leak Test

Target Pressure .................................................. 20 to 400 mmHg
Elapse time ......................................................... 0:30 to 5:00 minutes:seconds in 30 second steps
Leakage Rate ..................................................... 0 to 200 mmHg/minute
Internal Leak rate ................................................ <2 mmHg/min into 500 ml rigid volume

Pressure Relief Test Range ................................. 100 to 400 mmHg

Presets and Autosequences

Presets

Normal
Hypertensive
Hypotensive

Autosequences

Cardiac Failure sequence
Exercise sequence
Respiration sequence
Monitor testing sequence

Advertising