Invasive blood pressure simulation – Fluke Biomedical ProSim 4 User Manual

Page 28

Advertising
background image

ProSim™ 4

Users Manual

18

Note

When the battery discharges to a low level threshold, all NIBP
functions may be disabled.

Invasive Blood Pressure Simulation

The Product can simulate an invasive blood pressure transducer. To do an

invasive blood pressure simulation, connect the Product to the monitor as shown

in Figure 13.

ProSim 4

Patient

Monitor

IBP Cable

gne021.eps

Figure 13. Invasive Blood Pressure Connections

To access the invasive blood pressure parameters, touch the IBP control in the

Home screen to show the screen in Figure 14 in the display.

gne006.bmp

Figure 14. Invasive Blood Pressure Screen

The invasive blood pressure, heart rate, respiration, and the non-invasive blood

pressure parameters are shown at the top of the display. To change the invasive

blood pressure, touch one of the pre-defined blood pressure controls in the

display.
To zero the simulated output, touch the Zero softkey.
To set a static IBP pressure signal to the IBP monitor, touch the Static Pressure

softkey from the IBP screen. The static pressure screen in Figure 15 shows in

the display.