How to check a manometer – Fluke Biomedical ProSim 4 User Manual

Page 27

Advertising
background image

Vital Signs Simulator

Non-Invasive Blood Pressure Simulation and Tests

17

When the time expires, the leak rate is shown in the display. See Figure 11.

gne025.bmp

Figure 11. Leak Test Results Screen

Touch the Home softkey to go back to the Home screen.

How to Check a Manometer

To check a manometer, connect the Product to the manometer and manual

pump bulb as shown in Figure 12.

ProSim 4

Squeeze

Pressure Pump

Blood Pressure

Gauge

gne024.eps

Figure 12. Manometer Check Connections

1. Touch the More softkey.
2. Touch the Leak Test and Manometer control to show the screen shown in

Figure 10 in the display.

3. Squeeze the manual pump bulb to pressurize the line.
4. Check the value shown on the pressure gauge against the measurement in

the Product display.

Relieve the pressure through the pump relief valve before you remove the hose

from the Product.