Vital Signs Simulator

Autosequences

19

gne007.bmp

Figure 15. Static Pressure Screen

To set the simulated pressure, touch one of the pre-defined static pressure

controls.
Touch the Home softkey to go back to the Home screen.

Autosequences

Autosequences are a series of steps that change the output of the Product

automatically. For example, to do a monitor test, you could set normal ECG then

wait some time. Then touch Hypertensive and wait more time, then touch

Hypotensive and wait. The monitor testing autosequence does these changes

for you automatically. Each step of the monitor testing autosequence sets

simulation parameters and after the allotted time period, it does the subsequent

step.
To do an autosequence:
1. Touch the More softkey if it is shown in the Home screen.
2. Touch one of the autosequence controls to show the autosequence steps in

the display. The Monitor Testing autosequence is shown in Figure 16.

gne012.bmp

Figure 16. Monitor Testing Autosequence Screen