Ivoclar Vivadent UTS 3D User Manual

Page 15

Advertising
background image

15

This chapter describes the user maintenance and cleaning
procedure for the UTS 3D-transferbow. This list contains
only tasks that may be performed by dental personnel.
All other tasks must be performed by qualified service
personnel at a certified Ivoclar Vivadent Service Center.

7.1

Inspection and maintenance

The time for these maintenance procedures depends on
the frequency of use and the working habits of the users.
For that reason, the recommended times are only
approximate values.

7.2

Cleaning

Avoid contact with strong acids or solvents
(e.g. MMA). In this way, damage to the
surfaces can be prevented.

7.3

Cleaning instructions for the 3D-registration
joint

After every use or contact with saliva or blood,
wipe with a disinfectant.

Only use disinfectants that are suitable for stainless
steel and light metals.

Avoid any contact with strong acids, bases or
solvents.

Do not clean with ultrasonic bath, water, steam jet
or sterilization units.

What

Screws that are
contaminated
with wax or
plaster.

Screws that are
contaminated
with silicone.

3D-bite fork that
has come into
contact with
saliva or blood.

3D-registration
joint

UTS 3D-
transferbow

Nosepads

Earpieces

When

If required

If required

After every use

After every use

After every use

After every use

After every use

Cleaning material

Rinse with hot
water and dry
with a dry cloth

Clean with a
cloth.

see chapter 7.3

Clean with a
spray disinfectant

Clean with a
spray disinfectant

Use new nosepad
(nosepads are
disposable)

Autoclaving
device. Please
observe the
corresponding
instructions.

7. Maintenance, Cleaning, and Diagnosis