Maintenance, Clear tests, Clear alarms – Verilink PRISM 4101 (34-00230) Product Manual User Manual

Page 44: Loop, Unloop, Loop mode, Clear tests clear alarms loop unloop loop mode

Advertising
background image

32

C

HAPTER

4: T

ERMINAL

I

NTERFACE

Maintenance

When in-depth testing is necessary, you can use the Maintenance screen
(Figure 4-4) to perform loop tests and/or BERT functions on the DDS circuit.
BERT is performed by using on-board test facilities. Additional test equipment is
not required. Some of these tests may also be activated by the front panel buttons
as described in the section entitled Diagnostics on page 20.

Figure 4-4 Maintenance Screen

Clear Tests

Pressing Return clears all tests and any line loops that have been initiated.

Clear Alarms

Pressing Return clears all near end alarms. If alarm conditions remain present,
alarms are reactivated once thresholds are exceeded.

Loop

The type of loop is chosen by toggling the Spacebar and is executed by pressing
Return. Options are LOCAL, V.54, and FAR V.54.

Unloop

Pressing Return takes down the specified loop from the currently selected port.
The type of loop is chosen by toggling the Spacebar and is executed by pressing
Return. Options are LOCAL, V.54, and FAR V.54.

Loop Mode

Allows you to choose how the data will be looped back. The choices are

BIDIRECTIONAL or UNIDIRECTIONAL.

Advertising