Amprobe MEGATEST-1000 Megohmmeter User Manual

Page 12

Advertising
background image

5.1.2.1 "AUTO" MODE

CAL

LOW

1.07

219

mA

AUTO

At the end of the test, if the average
resistance value Ravg
results to be
lower than 5

Ω the instrument emits a

double sound signal indicating the
positive outcome of the test and
displays one screen similar to the
screen alongside.

Average test
current value
Iavg.

Average
resistance
value Ravg.


SAVE

The test can be stored pressing the SAVE key twice (according to
paragraph 0).



5.1.2.2 "R+" or “R-" MODE

CAL LOW

1.07

219

mA

R+
R-

At the end of the test if a resistance
value R+ or
R- lower than 5

Ω was detected, the

instrument emits a double sound
signal
indicating

the positive

outcome of the test and displays
one screen similar to the screen
alongside.


Resistance
value R+ or
R-.

Value of test
current I+ or I-.

Either the symbol R

+

or

the symbol R

-.


SAVE

The test can be stored pressing the SAVE key twice (according to
paragraph 0).



5.1.2.3 "R+TIMER" or “R-TIMER" MODE

CAL LOW

1.08

219

mA

5

s

R+
R-

TIMER

At the end of the test if a resistance
value R+Timer
or R–Timer lower
than 5

was detected, the instrument

(after the set time has elapsed) emits
a double sound signal indicating
the positive outcome
of the test
displays one screen similar to the
screen alongside.

Duration time of
the test.

Value of the
test current I+
or I-.

The symbols R+ or R- are displayed.




SAVE The test can be stored by pressing the SAVE key twice (see paragraph 0).

11