The following symbols are used in this manual:

Caution: refer to the instructions reported in this manual; improper use may
damage the apparatus or its components.



Rotary switch of the instrument.

2.3 DURING

USE


Carefully read the following recommendations and instructions:

o

WARNING:

Non compliance with the Warnings and/or Instructions may
damage the instrument and/or its components or injure the
operator.

Before selecting any function disconnect the test leads from the circuit under test.

When the instrument is connected to the tested circuit never touch any test lead which
is not being used.

Avoid taking resistance measurements in the presence of external voltages; even
though the instrument is protected, high voltage may cause malfunctions.

o

WARNING:

If the symbol

is displayed during use interrupt testing and

replace batteries following the procedure described under
paragraph 9.2.

Note: The instrument is capable of keeping the data stored even

though batteries are not installed.


2.4 AFTER

USE

After the measurements are completed turn the rotary switch to the “OFF” position.

Remove batteries if the instrument is to remain unused for long periods of time.