5.3.3 CLEAR

If you want to cancel the stored tests results proceed as follows:


1. Press

the

RCL key. The instrument displays a screen like the

following:

RCL

MEM

3

P02

Number of the memory location in which
the measurement results were stored.

Value of the parameter P related to the
saved measurement.


2.

Use the keys ▼, ▲ to select the number of the memory location.


ATTENTION: the instrument will cancel all the results stored from the

memory location selected to the last memory location
full.


3. Press

the

DISP key to display the test result related to the memory

location selected. In case that a few results were stored in the memory
location (Ex: RCD test under AUTO mode) it is necessary to press
DISP more than once to display them.

DISP


4.

Press CLR once. The blinking symbol "

clr"

is displayed. Now you

face two possibilities:

CLR

MEM

clr

p 08

The instrument cancels the memory cells from n.2
to n.8, where n.2 is the one selected by the
operator while n.8 is the last test saved in the
instrument.

Blinking.

CLR

Press CLR again if the tests selected are to be cancelled starting
from the one selected down to the last one saved.

ESC

Press ESC to nullify the clearing phase. The blinking symbol "clr"
disappears.


ESC

Press the ESC key at any point if you want to quit the memory management
mode and return to the selected measurement function.