Reset procedure, Default parameters, Default parameters low – Amprobe MEGATEST-1000 Megohmmeter User Manual

Page 25: Default parameters m, Memory default parameters

Advertising
background image

Example:

97 tests have been stored in the instrument.

If you want to cancel the tests from 43rd to 97

th

, you proceed as follows:

- Press RCL.
- Select with keys ▼, the memory location 43.
- Press CLR. You can see CLR blinking on the primary display, 43 on the

left secondary display and 97 on the righ secondary display.

- Press CLR. The tests from 43 to 97 has cancelled.

5.4 RESET

PROCEDURE


Reset of the INSTRUMENT and default parameters

Press DISP, CLR, RCL at the same time the rotary switch is turned on.

rES

The screen nearby is displayed
for 5 seconds
, after which the
instrument emits a sound signal
and then displays the screen
relative to the function selected
with the rotary switch.


ATTENTION: the reset procedure will erase all the stored tests and it will set the default

parameters in the instrument.


Before conducting a reset procedure, download the stored tests on a PC.

5.5 DEFAULT

PARAMETERS


5.5.1 DEFAULT PARAMETERS LOW

Parameter

Reset default parameter

Mode AUTO
Calibration Offset

0 (CAL off)

Mode R+/R- TIMER

Timer 3s



5.5.2 DEFAULT PARAMETERS M

Parameter

Reset default parameter

Mode MAN
Test voltage

500V

Mode TIMER

Timer 60s



5.5.3 MEMORY DEFAULT PARAMETERS

Parameter

Reset default parameter

Parameter P

P = 1

Memory state

1(empty)

24