Magnetron removal/replacement – Fusion F300S User Manual

Page 89

Advertising
background image

Installation, Operation, and Maintenance Manual # 509252 Rev R

89

5. Reconnect the filament transformer wires to the magnetron and tighten the

terminal screws to 0.9 N-m (8 inch-lbs) using a torque-limiting screwdriver.
Reposition the tubing over the terminal screws and install cable ties.

6. Install the irradiator housing as described in

“I300M Irradiator Housing

Removal/Replacement” on page 86

.

Magnetron Removal/Replacement

1. Remove the irradiator housing as described previously.

2. Disconnect filament transformer wires from magnetron by removing the cable

ties and sliding back the tubing. Remove the terminal screws.

3. Remove the filament transformer bracket and air plenum.

4. Remove the four screws fastening the magnetron to the casting. Lift the

magnetron off the waveguide assembly and place it upside down in a clear area
on a wooden bench, at least five inches from any magnetic materials (such as
iron or steel).

5. Install the replacement magnetron and fasten it to the casting using the

associated hardware.

6. Refer to

Figure 29

; hand tighten the magnetron fasteners in sequence as shown

until snug. Then, tighten the fasteners ¼ turn in sequence. Finally, tighten the
fasteners in sequence to 3.4 N-m (30 inch-lbs) using a torque-controlled driver.

Figure 29. Magnetron Fastener Tightening Sequence (I300M)

7. Install the filament transformer bracket and air plenum.

8. Reconnect the filament transformer wires to the magnetron and tighten the

terminal screws to 0.9 N-m (8 inch-lbs) using a torque-limiting screwdriver.
Reposition the tubing over the terminal screws and install cable ties.

9. Install the irradiator housing as described in

“I300M Irradiator Housing

Removal/Replacement” on page 86

.

Advertising