Glucose Control Solutions

Use SureStepPro Glucose Control Solutions
with the SureStepPro Bedside Unit and
SureStepPro Test Strips only.

IMPORTANT Control Solution Information

Read the SureStepPro Control Solutions
package insert for more information.

• Gently shake each vial of control solution

before use.

• Do not use a control solution after the

expiration date printed on the vial label.

• Discard any unused portion 3 months after

opening. Write the opened date on the vial
when you first open it.

• The control ranges printed on the

SureStepPro Test Strip bottle label are for
SureStepPro Control Solutions only.

• Store the control solution at room

temperature below 30°C (86°F). Do not
refrigerate or freeze.

• The control solutions contain dye that

stains clothing.

• The control solutions are not for human

consumption.

• Control solution values can be affected at

altitudes of 3000 feet or greater. Read the
SureStepPro Glucose Control Solutions
package insert for specific information and
correction factors.