Body, Troubleshooting – Lifescan SureStep Pro User Manual

Page 96

Advertising
background image

Troubleshooting

84

If you cannot resolve a problem, call LifeScan Healthcare Professional Line:

1 800 524-7226 (USA)

1 888 353-0800 (Canada)

Error 6
System Error

The bedside unit
detected a problem
with the electronics.

1.Wipe contact points in lens

area with cloth moistened
with water to remove
residual bleach. Dry
thoroughly, but
NOTE: Before cleaning
contact points, compare
color of test strip confirma-
tion dot to Color Chart on
test strip bottle. If glucose
level is very high or very low,
take appropriate steps to
monitor patient and provide
medical care.
• Repeat test with new
test strip.
• Make sure test strip lot
number (and code)
entered at bedside unit
matches number on test
strip bottle.

2.If you continue to get any

error message, obtain a
laboratory result. Take
appropriate steps to
monitor patient and
provide medical care while
awaiting laboratory result.

3.Contact LifeScan for help.

ERROR

POSSIBLE

CAUSE

SOLUTION

BU.book Page 84 Monday, October 28, 2002 11:14 PM

Advertising