Troubleshooting

82

If you cannot resolve a problem, call LifeScan Healthcare Professional Line:

1 800 524-7226 (USA)

1 888 353-0800 (Canada)

Error 4
Temperature
Range

The bedside unit
detected a
temperature
problem.

1.Repeat test with new test

strip. Make sure bedside
unit and test strip are at
room temperature.
• Do not apply heat or
cold to unit (operating
temperature is 18°–30°C).

2.If you continue to get any

error message, obtain a
laboratory result, and
• Compare color of test
strip confirmation dot to
Color Chart on test strip
bottle to see if glucose
level is very high or very
low.
• Take appropriate steps
to monitor patient and
provide medical care while
awaiting laboratory result.

3.Contact LifeScan for help.

ERROR

POSSIBLE

CAUSE

SOLUTION