Patient Test

36

Check the white pad on the front of the test
strip and the confirmation dot on the back of
the test strip.

• If the white pad becomes completely

saturated, you have applied too much blood
for an accurate result. Repeat the application
with a new test strip.

• The confirmation dot on the back of the test

strip should be completely blue.

If white patches or streaks are visible on the
blue confirmation dot, you may not have
applied enough blood for an accurate result.
Repeat the application with a new test strip.

8

FRONT

confirmation
dot

BACK

FRONT

BACK