VEGA VEGAPULS 51V…54V User Manual

Page 20

Advertising
background image

20

VEGAPULS 51 V … 54 V

Ex-technical data

(note approval documents)

Flame proofing

ia (intrinsically safe in conjunction with
a safety barrier or separator)

Classification

EEx ia IIC T6

Perm. housing ambient temperature

T6: 45°C

T5: 58°C

T4/T3: 60°C

Temperature class (permissible ambient temperature on the antenna system)

- T6

80°C

- T5

95°C; type 51: 80°C

- T4

130°C; type 51: 80°C

- T3

150°C; type 51: 80°C; type 52: 130°C

Ex-approved in category or zone

- EC-type approval

Type 5*V Ex: Zone 1 (II 2 G)
Type 5*V Ex 0: Zone 0 (II 1 G)

- conformity certificate

Type 5*V Ex: Zone 1
Type 5*V Ex 0: Zone 0

Process connections

VEGAPULS 51, 52

G 1

1

/

2

A, 1

1

/

2

" NPT (rod antennas on plastic

or StSt thread)

VEGAPULS 53

DN 50, DN 80, DN 100, DN 150 (rod antennas)

VEGAPULS 54

DN 50, DN 80, DN 100, DN150
ANSI 2", 3", 4" and 6"
(up to DN 100 or 4" mounting on surge pipe)

Connection lines

Two-wire sensors, supply and signal via one two-wire line
Line resistance max. 15

Ω per wire or 1000 m cable length

Cross-section area of conductor

generally 2,5 mm

2

Earth connection

max. 4 mm

2

Cable entry

2 x M20 x 1,5 (cable diameter 5 … 9 mm)

Materials

Housing

PBT (Valox)

Flange / process connection

- VEGAPULS 51

PVDF or StSt

- VEGAPULS 52

PVDF or StSt

- VEGAPULS 53, 54

1.4571

Antenna

- VEGAPULS 51

PP or StSt / PPS

- VEGAPULS 52

PTFE or StSt / PTFE

- VEGAPULS 53

PTFE

- VEGAPULS 54

1.4571, 1.4071

Flange coating (only VEGAPULS 53)

PTFE

Technical data

Advertising