38

VEGAMET 514

Settings in the "extended menu"

7.7 Conditioning

The menu items "

Scaling

" and "

Integration

time

" correspond to the menu items in the

"reduced menu" (see appropriate chapter).

The menu items "

Density…

" or "

DK value…

"

are optionally displayed when under "Con-
figuration measurement loop" the appropriate
sensor type has been selected. The density
or DK value must, before adjustment with
medium, be so adjusted that it corresponds
to the value of the medium (see also chapter
"7.6 Adjustment").

The conditioning of VEGAMET can be
adapted to different vessel forms. The follow-
ing linearisation curves are available:
- linear (factory setting)
- horiz. cylindrical tank
- spherical tank
- Lin-curve 1
- Lin-curve 2
- Lin-curve 3

Lin-curves 1 … 3 can be edited by the
user (see appropriate chapter).

After a scaling or linearisation, the required
output must be assigned to the scaled or
linearised value (menu item "prop. to..." and
"Unit...") under the menu "Parameter adjust-
ment outputs".

Under

Meas. value limitation

you can ex-

clude negative values.

Failure mode Standard

means that at a sensor current of 3.6 mA or
21 mA the fail-safe relay responds.

Failure mode >110%

also generates a fault signal when a meas.
value of more than 110 % or less than -10 %
is reached.

Param.
TAG-
No. 1

Adjust-
ment

Signal
condi-
tioning

TAG -
No. 1
%
xx,x

Scale

Lin.
curve

linear

0%
corres-
ponds

0

100%
corres-
ponds

1000

Integr
ation
time s

0

Density
in
Kg/dm

3

1,000

Meas.
value
limita-
tion

Deci-
mal
point

888,8

prop.
to

unde-
fined

Unit

– –

nega-
tive
values

yes

Failure
mode

Stan
dard

DK
value
(Er)

81,00

cylindrical tank
etc.

>110%