VEGA VEGAMET 514 User Manual

Page 7

Advertising
background image

VEGAMET 514

7

Voltage output

Number

1 output

Function

analogue output of the processing results

Range

adjustable between 0 … 10 V

Current

max. 1 mA

Resolution

0.5 mV

Linearity error

0.05 % (relating to 10 V)

Temperature error

0.06 %/10 K (relating to 10 V)

Relay output

Number

2 switching relays
1 fail-safe relay

Contact

floating spdt

Contact material

AgNi, hard gold plated

Turn-on voltage

min. 10 mV DC
max. 250 V AC/DC

Switching current

min. 10 µA
max. 3 A AC, 1 A DC

Breaking capacitance

max. 500 VA, 54 W

Min. switching hysteresis
(Low-/High delta)

0.5 %

DISBUS output

Function

for linking the signal conditioning instruments

and for connection of digital indicating
instruments

Connection cable

2-wire standard cable (screening is
recommended)

max. cable length

1000 m

Indicating elements

Clear text indication

LC display

(not available with VEGAMET 514N)

- 4-line, 6 digits each
- background lighting

Analogue indication

LED chain consisting of:

(not available with VEGAMET 514N)

- 11 segments 0 % … 100 %
- shows the actual value of the measurement

loop

LEDs in front plate

green: operating voltage on
red: fault signal (LED lights with deenergised
relay)
yellow: relay status (standard setting:
LED lights with energised relay)

Adjustment elements

Front plate

6 keys for configuration and parameter
adjustment
(not available with VEGAMET 514N)

Circuit board

rotary switch for adjustment of the instrument

address on DISBUS

Product description

Advertising