2 adjustment with pc – VEGA VEGAPULS 81 User Manual

Page 55

Advertising
background image

VEGAPULS 64 and 81

55

Note

A digital adjustment signal is superimposed to

the 0 … 20 mA signal output. Is the inner

resistor of the connected signal conditioning

instrument is less than 100

(e.g. with an

indicating instrument), the digital adjustment
signal is considerably damped. In this case

loop a load resistor R

100

into the signal

line to avoid the low impedance damping

during the adjustment.

Adjustment

The PC with the adjustment software VVO is

connected to the measuring system.

The individual adjustment and input steps are

described in the following (e.g. • choose…; •
Start…).

• Switch on the power supply of the con-

nected radar sensor.

• Switch on the power supply of the con-

nected signal conditioning instrument.

• Start your PC and the adjustment software

VVO.

• In the entrance screen choose with the shift

keys or with the mouse the point “

Planning“

and push “

OK“. The identification is en-

quired.

Set-up

6.2 Adjustment with PC

Connection

In the chapter “2.3 Configuration of measuring
systems“ the connection of the PC is shown in

different co-ordinations. The PC with the ad-

justment program VVO (VEGA Visual Operat-

ing) can be connected to the:

- sensor
- signal line

- signal conditioning instrument VEGAMET

514 V or 515 V

- processing system VEGALOG

Connection of the PC to a sensor

For connection of the PC to a sensor,

VEGACONNECT interface converter is re-

quired. Plug the VEGACONNECT into the
standard connection of the PC and connect

the two-wire line of VEGACONNECT either to

the signal output on the sensor (VEGAPULS

81) or (for instruments of VEGAPULS 64 se-

ries) to the provided CONNECT-socket in the
sensor.

Connection of the PC to a signal line

Connect the two-wire line of VEGACONNECT

to the signal output line of the sensors.

Connection of the PC to the signal condi-
tioning instrument VEGAMET

Plug the two-wire line of VEGACONNECT into

the CONNECT-sockets on the signal condition-

ing instrument.

Connection of the PC to the processing
system VEGALOG

Connect the PC with the processing system

VEGALOG 571 via a standard computer cable.

• Enter under the name “

VEGA“.

• Enter under the password “

VEGA“.

This is the pre-adjusted identification.

Advertising