VEGA VEGAPULS 81 User Manual

Page 59

Advertising
background image

VEGAPULS 64 and 81

59

Set-up: Adjustment with PC

Parameter adjustment / Adjustment

• Click to the button “

Adjustment“.

The window “

Adjustment“ appears.

• Choose the sensor and confirm with “OK“.

The main menu “

Instrument data parameter

adjustment“ appears.

• Click the button “

Min / Max-Adjustment“.

Choose the following window if you want to

carry out the empty and full adjustment with or

without measured product.

When you chose the adjustment with meas-

ured product, you have to inform the sensor,

which level or which distance corresponds to

0% and 100%. If for full you have e.g. just filled
up to 94%, you can correct the percentage

value accordingly from 100% to 94%.

Choose adjustment without measured product

and you can determine what should be full
and what should be empty.

• Choose “

no (adjustment without measured

product)“.

The following picture is a distance measure-

ment, 0 % correspond to a distance to the

measured surface of 0.000 m. 100 % to a

distance of 6.000 m.

The measuring window of the sensor is hence

0 … 6 m.

• Cho0se the distance which corresponds in

your application to 0 % and 100 %.

Advertising