VEGA VEGAPULS 81 User Manual
Page 59
VEGAPULS 64 and 81
59
Set-up: Adjustment with PC
Parameter adjustment / Adjustment
• Click to the button “
Adjustment“.
The window “
Adjustment“ appears.
• Choose the sensor and confirm with “OK“.
The main menu “
Instrument data parameter
adjustment“ appears.
• Click the button “
Min / Max-Adjustment“.
Choose the following window if you want to
carry out the empty and full adjustment with or
without measured product.
When you chose the adjustment with meas-
ured product, you have to inform the sensor,
which level or which distance corresponds to
0% and 100%. If for full you have e.g. just filled
up to 94%, you can correct the percentage
value accordingly from 100% to 94%.
Choose adjustment without measured product
and you can determine what should be full
and what should be empty.
• Choose “
no (adjustment without measured
product)“.
The following picture is a distance measure-
ment, 0 % correspond to a distance to the
measured surface of 0.000 m. 100 % to a
distance of 6.000 m.
The measuring window of the sensor is hence
0 … 6 m.
• Cho0se the distance which corresponds in
your application to 0 % and 100 %.