Appendix A. Pop-up and Error Messages

Alarms

Sensor Error

Alarm - The converter has expe-
rienced HW problems measuring
flow (see log for details/time)

Sensor cal. Code error

Alarm - The calibration code con-
flicts with the sensor HW.

Output Coil Error

Alarm - The converter has experi-
enced problems with connection to
the sensor coil (see log for details/
time)

Empty pipe

Alarm - The converter has detect-
ed no water in the sensor

Input over range

Alarm - Sensor flow can not be
measured correct. Input is over
max. level. Will normally only occur
starting from zero flow to high flow.
Will be present for max. 12 min-
utes.

Factory reset has occurred

Alarm - A factory reset has oc-
curred (some settings have
changed, check all used settings)

Eeprom write error

Alarm - Eeprom write error (see log
for details)

Flow total counter eeprom error

Alarm - Flow totalize Eeprom error
(non- resettable counter maybe
incorrect in Eeprom)

Sensor Setting error

Alarm - Sensor Setting error (some
settings are incorrect or changed)

Counter output Forward Pulsover-
flow

Alarm - Counter output Forward
Pulse overflow (too low pulse
volume)