Electromagnetic Flow Meter

FW: 842018/842009

Wiring Schematic #1

for "Remote Converter with Connection Box on Flow Sensor" on the op-
posite page (page

34

).

Flow Sensor
Connect one end of
the MJK cable to the
flow sensor as de-
scribed in the table
below. All 3 shields are
twisted and connected
in slot 1. The white
lead in pair no. 3 is not
used.

B Connection for remote converter

Terminal Signal

Wire

Colour before

March 2001

Colour from
March 2011

1

GND

Shield

-

-

2

E1

Pair no. 1

Red

Red

3

E2

Pair no. 1

White

White

4

GND

Pair no. 3

Black

Black

-

-

Pair no. 3

White

-

5

L1

Pair no. 2

Green

Blue

6

L2

Pair no. 2

White

Brown

7, 8, 9

not used

Remote Converter
(and Display Unit)
Connect the other end
of the MJK cable to
the remote converter
as shown to the right.
The "1 Shield" lead
goes into "Ground
from flow pipe", the
"2 Red" lead goes
into "E1 Red", the "3
White" lead into "E2
White", and so on.