Digital input local (compact) flow sensor, Digital input, Local (compact) flow sensor – Xylem Magflux Electromagnetic Flow Meter User Manual

Page 30: Magflux

Advertising
background image

30

EN 3.05 MagFlux Manual 1212

Electromagnetic Flow Meter

See also application examples in Appendix G. Digital Input/Output Con-
nections on page

123

.

Digital Input

MagFlux

®

has one digital input which is activated with a voltage higher

than 10 V DC and de-activated with a voltage lower than 5 V DC.

Digital Input

Terminal

Designation

DI

Com

DI

Max. 30 V DC

The digital input (DI) can be programmed for the following functions:

-

start and pause batch 1 and batch 2 counters

-

reset counters R Totalizer Forward and Reverse

-

reset counter R Totalizer Sum

DI shares the common terminal (Com) with DO 1.
See also application examples in Appendix G. Digital Input/Output Con-
nections on page

123

.

Local (compact) Flow Sensor

Connect the compact (local) flow sensor to the flow converter with the
wires coming from the flow sensor as shown below.

Compact Flow Sensor

Terminal

Signal

Designation

Colour from sensor

1

Liquid GND

Built-in liquid
ground electrode

Black/shield

2

E1

Electrode

Red

3

E2

Electrode

White

4

GND

Ground

Black

5

L1

Coil

Blue

6

L2

Coil

Brown

7, 8 and 9

not used