Electromagnetic Flow Meter

FW: 842018/842009

Totalizers

The MagFlux

®

provides six totalizers, each with two or three output op-

tions. Up to two totalizers can be assigned to one digital ouput.
The totalizers are:

RFTot

Resettable Forward Totalizer

RRTot

Resettable Reverse Totalizer

RSTot

Resettable Sum Totalizer

and
NR FTot

Non-Resettable Forward Totalizer

NR RTot Non-Resettable Reverse Totalizer
NR Cnt

Non-Resettable (Counter) Sum Totalizer

The available output options for the resettable totalizers are:

Mechanical Relay (DO2)
Opto Relay (DO1) (light-triggered electronic relay)
Display only (on the Display Unit)

The available output option for the non-resettable totalizers is:

Display only (on the Display Unit)

If a chosen output is not available (dedicated to another function), a
"Device Exception" pop-up message will inform the operator.

Settings and Limits for Resettable Totalizers

Totalizer Forward Unit

Available units are l, hl, kl, m

3

, ft

3

, gal and MG.

Reset with Digital Input

Reset the totalizer via the digital input (DI)

On-time

The pulse length of the opto digital output (DO 1) is adjustable from 1
ms to 10 secs.
The pulse length of the mechanical digital relay (DO 2) is adjustable
from 100 ms to 10 secs.