69

Electromagnetic Flow Meter

FW: 842018/842009

If a required volume betwen pulses is not contained in the above steps,
for example for a pre-set to a sampler, use the batch counter option.
The relays are limited to a minimum and a maximum of pulses pr. hour
depending on the 20 mA scale and the on-time of the relay in milli-
seconds.

DO 1 (Opto Relay)

The maximum limit is 500 pulses pr sec at 1 ms pulse length.

The formula for the minimum flow unit pr pulses is:

(flow at 20mA) x (pulse delay in mSec) / 1800000

Example: Min. flow unit/pulses for DN 100 max. flow 300 m

3

/h ?

Minimum flow unit pr pulses = 300 m

3

x 100 mSec / 1800000 =

0,016 m

3

which is rounded to 0,1 m

3

by the MagFlux.

The formula for the maximum flow unit pr pulses is:

(flow at 20mA)

Example: DN 100 max. flow 300 m

3

/h ?

Maximum flow unit pr pulses = 300 m

3

DO 2 (Mechanical Relay)

The maximum limit is 120 pulses pr hour.

The formula for minimum flow unit pr pulses is:

(flow at 20mA) / 120

Example: DN 100 max. flow 300 m

3

/h ?

Minimum flow unit pr pulses = 300 / 120 = 2,5 m

3

which is rounded to

10 m

3

by the MagFlux.

The formula for maximum flow unit pr pulses is:

(flow at 20mA)

Example: DN 100 max. flow 300 m3/h ?
Maximum flow unit pr pulses = 300 m

3