The "Service Menu" provides options intended for service personnel dur-
ing installation, calibration, operation, monitoring and maintenance.
All parameters can be read without any restrictions, but certain parameters
can only by changed after a password has been typed.

Flow Direction

Mechanical Relay (DO2)
Opto Relay (DO1)
Display only
Not in use

Back

Select

OK

Mechanical Relay (DO2)

Opto Relay (DO1)

Display only

Relay Function
NO
NC

Back

Select

OK

Empty Pipe

Mechanical Relay (DO2)
Opto Relay (DO1)
Display only
Not in use

Back

Select

OK

Delay

00010 Sec

Back

Select

>

Delay

00010 Sec

Back

Select

>

Delay

00010 Sec

Back

Select

>

Mechanical Relay (DO2)

Opto Relay (DO1)

Display only

Relay Function
NO
NC

Back

Select

OK

Delay

00010 Sec

Back

Select

>

Delay

00010 Sec

Back

Select

>

Delay

00010 Sec

Back

Select

>

Type Password

0000

Back

Select

>

Service Menu

Sensor Calibration Code
Converter SW Ver.
Product Data
Reset Counter Time
Internal Meas. & Cal.
Minimum Velocity
Calibrate Flow now
Calibrate Volume
Calibrate mA
Test Velocity
Freeze Coil
Read Event Log
Back on Stock
Back

Select

OK

OK

Only f

or ter

minal use

Must not be surrended

A

B

C

D

E

Date

Modification

Scale:

MJK AUTOMATION A/S
Byageren 7
DK-2800 Nñrum
Fax: +45 45 56 06 46
Tel.: +45 45 56 06 56

Date:

Constructor:

Part no.:

Software version

Skala for tegning

Menu system for MagFlux Converter
Konverter Setup

Service Menu, Flow Direction, Empty Pipe

2006 - 09 - 20

841010

KJ/LSP

Path: Main Menu -> Converter setup -> Flow Direction / Empty Pipe / Service Menu / Status

Relay Function
NO
NC

Back

Select

>

Relay Function
NO
NC

Back

Select

>

Status

DO1

Not in use

DO2

R Totalizer Forward

DI

R Totalizer Forward

mA output 6.250 mA

Back

Select

OK

Sensor Calibration Code
Each and every MagFlux

®

has a unique calibration code that contains

calibration data for the specific unit. A calibration code could for exam-
ple be dw5uq4 or 7v3ri0.
The current converter firmware requires an 8-character input, but it also
accepts 6 characters plus two "Ok"s. If the code is dw5uq4, enter:

d w 5 u q 4 "Ok" "Ok"

where "Ok" indicates that you press the "Ok" key without selecting any
character. Only small letters can be entered from the MagFlux

®

key-

pad.

Advertising