Luminex 100 User Manual Version 1.7 User Manual

Page 120

Advertising
background image

Luminex 100 User Manual Version 1.7

x

MAP Technology

9 - 10

PN 89-00002-00-063 Rev. A

Technical notes

1. Oligonucleotides must be synthesized with a 5’ or 3’

amino

group. No Tris or Azide or other amine-containing buffers
should be present during the coupling procedure. Preferably, the
oligonucleotide should be resuspended in water following
synthesis. If oligonucleotides were previously solubilized in an
amine-containing buffer, then precipitation and resuspension into
water is required.

2. This procedure can be scaled up or down.

3. The optimal coupling concentration for a given oligonucleotide

is determined by coupling at various concentrations within the
recommended range of 1 to 50 µM.

4. Minimize the exposure of EDC to air; secure closures on stock

and dry aliquot containers, store desiccated at -20°C. Use
aliquots immediately and discard containers after use. Make a
fresh 10 mg/ml EDC solution before each addition.

Binding Biotin-

conjugated

Molecules to

LumAvidin-Modified

xMAP Microspheres

Introduction

Use the protocols included here as a general starting point for
developing assays. All assays should be optimized for your reagents
in your specific application. You will discover the methods that give
you the best results by starting with these guidelines and modifying
them for your specific needs.

Equipment

Vortex

Micropipetters (1 µl - 1000 µl)

Microcentrifuge

Timer

Rotator

Materials

xMAP LumAvidin-modified microspheres—LIMIT
EXPOSURE TO LIGHT!

Biotin-conjugated molecule

Microcentrifuge tubes: 1.5 ml, polypropylene