Maintenance and Cleaning

PN 89-00002-00-063 Rev. A

7 - 7

Clean Luminex 100

Analyzer Ventilation

Filter

‹

Check the Luminex 100 analyzer ventilation filter every 6

months. Clean the filter only when soiled.

1. Turn off the Luminex 100 analyzer at the power switch.

2. Unplug the power cord from the wall source.

3. On the bottom of the unit on the left side, push the filter out

toward the right side.

4. Grasp the filter and pull it out from underneath the Luminex 100

analyzer.

5. If the filter is dirty, clean it with a vacuum cleaner or in running,

distilled water. Stand it upright to air-dry.

6. Reinstall the filter with the arrows facing up.

Change Sheath

Filter

‹

Change the sheath filter once a year.

1. Disconnect the sheath fluid bottle

2. Squeeze the metal quick release connections at both ends of the

filter assembly, then pull the tubing away from the filter
assembly.

3. Connect the new sheath filter, matching up the color-coded

fittings. The arrow on the sheath filter should point up.

4. Reconnect the sheath fluid bottle.

5. Prime the instrument 5 times to remove any air introduced into

the system.

Figure 7-5. Sheath Fluid Filter