Teledyne 3010TB - Split architecture trace oxygen analyzer User Manual

Page 23

Advertising
background image

Oxygen Analyzer

Part I: Control Unit

Part I: 3-7

Teledyne Analytical Instruments

There are three alarm-circuit connectors on the alarm relays block

(under RELAY OUTPUTS) for making connections to internal alarm relay
contacts. Each provides a set of Form C contacts for each type of alarm.
Each has both normally open and normally closed contact connections. The
contact connections are indicated by diagrams on the rear panel. They are
capable of switching up to 3 ampers at 250 V AC into a resistive load
(Figure 3-6).

Figure 3-5: Types of Relay Contacts

The connectors are:

Threshold Alarm 1:

• Can be configured as high (actuates when
concentration is above threshold), or low
(actuates when concentration is below thresh old).
• Can be configured as fail-safe or non-fail-safe.
• Can be configured as latching or nonlatching.
• Can be configured out (defeated).

Threshold Alarm 2:

• Can be configured as high (actuates when concen-

tration is above threshold), or low (actuates when
concentration is below threshold).

• Can be configured as fail-safe or non-fail-safe.

• Can be configured as latching or nonlatching.
• Can be configured out (defeated).

System Alarm:

Actuates when DC power supplied to circuits is

unacceptable in one or more parameters. Permanently
configured as fail-safe and latching. Cannot be de-
feated. Actuates if self test fails.

To reset a System Alarm during installation, discon-

nect power to the instrument and then reconnect it

Further detail can be found in chapter 4, section 4-5.

Advertising