Teledyne 3010TB - Split architecture trace oxygen analyzer User Manual

Page 41

Advertising
background image

Oxygen Analyzer

Part I: Control Unit

Part I: 4-13

Teledyne Analytical Instruments

4.4.1.2

Manual Mode Zeroing

Press

Zero

to enter the

Zero

function. The screen that appears allows

you to select between automatic or manual zero calibration. Use the

Δ∇ keys

to toggle between AUTO and MAN zero settling. Stop when MAN appears,
blinking, on the display.

Zero: Settling: Man
<ENT> To Begin

Press

Enter

to begin the zero calibration. After a few seconds the first

of five zeroing screens appears. The number in the upper left hand corner is
the first-stage zero offset. The microprocessor samples the output at a prede-
termined rate. It calculates the differences between successive samplings and
displays the rate of change as Slope= a value in parts per million per second
(ppm/s).

#### ppm Zero
Slope=#### ppm/s

NOTE:It takes several seconds for the true

Slope

value to display.

Wait about 10 seconds. Then, wait until

Slope

is sufficiently

close to zero before pressing Enter to finish zeroing.

Slope

is

given in ppm/s.

Generally, you have a good zero when Slope is less than 0.05 ppm/s

for about 30 seconds. When Slope is close enough to zero, press

Enter

. In a

few seconds, the screen will update.

Once span settling completes, the information is stored in the

microprocessor, and the instrument automatically returns to the

Analyze

mode.

4.4.1.3

Cell Failure

Cell failure in the 3010TB is usually associated with inability to zero

the instrument down to a satisfactorily low ppm reading, e.g. cell does not
fail if it comes below 5PPM . When this occurs, the 3010TB system alarm
trips, and the LCD displays a failure message.

#.# ppm Anlz
CELL FAIL/ ZERO HIGH

Before replacing the cell:

a. Check your span gas to make sure it is within specifications.

b. Check for leaks downstream from the cell, where oxygen may be

leaking into the system.

Advertising