Teledyne 3010TB - Split architecture trace oxygen analyzer User Manual

Page 46

Advertising
background image

4 Operation

Model 3010TB

4-18: Part I

Teledyne Analytical Instruments

AL—1 AL—2
Choose Alarm

Set up alarm 1 by moving the blinking over to AL–1 using the < >

arrow keys. Then press

Enter

to move to the next screen.

AL—1 1000 ppm HI
Dft—N Fs—N Ltch—N

Five parameters can be changed on this screen:

Value of the alarm setpoint, AL–1 #### (ppm or % oxygen)

Out-of-range direction, HI or LO

Defeated? Dft–Y/N (Yes/No)

Failsafe? Fs–Y/N (Yes/No)

Latching? Ltch–Y/N (Yes/No).

To define the setpoint, use the < > arrow keys to move the
blinking over to AL–1 ####. Then use the

Δ∇ arrow keys to

change the number. Holding down the key speeds up the
incrementing or decrementing. (Remember, setpoint units are
parts-per-million.)

To set the other parameters use the < > arrow keys to move the
blinking over to the desired parameter. Then use the

Δ∇ arrow

keys to change the parameter.

Once the parameters for alarm 1 have been set, press

Alarms

again, and repeat this procedure for alarm 2 (AL–2).

To reset a latched alarm, go to Dft– and then press either

Δ two

times or

∇ two times. (Toggle it to Y and then back to N.)

–OR –

Go to Ltch– and then press either

Δ two times or ∇ two times.

(Toggle it to N and back to Y.)

4.6

The Range Function

The Range function allows the operator to program up to three concen-

tration ranges to correlate with the DC analog outputs. If no ranges are
defined by the user, the instrument defaults to:

Range

Limits

Low

0–100 ppm

Med

0–1,000 ppm

High

0–10,000 ppm.

Advertising